Avante Velocity EKG 12-channel PC-based EKG
DETAILS
The Avante Velocity EKG is an affordable, all-in-one EKG solution.
The Avante Velocity EKG is a 12-channel, PC-based EKG that provides clinicians with an affordable solution for both resting ECG and stress test measurement. With the high-quality acquisition module and accompanying software, you can record, analyze, store and review 12-channel ECG’s.
12-Channel ECG
The Avante Velocity EKG offers clear and reliable ECG waveform acquisition tested with CSE/MIT/AHA databases.
Comprehensive Analysis
Use the Avante Velocity software to analyze optional advanced parameters including: Vector ECG, Timed VCG, Signal Average Electrocardiogram (SAECG), Heart Rate Turbulence (HRT), QT Interval Dispersion, High Frequency ECG, and Frequency Spectrum Cardiogram.
Stress Test Capabilities
The Avante Velocity EKG is compatible with industry-leading brands of treadmills and bicycles and supports many exercise blood pressure monitors. Treadmill and monitor sold separately.
Features
Simultaneous 12-lead acquisition
Real-time EKG waveform display on PC screen
Auto/manual measurement and interpretation
Large data storage and powerful data management software
Reports in Word, PDF and JPG formats — easily transfer via email
30s – 360s HRV analysis
Stress Test (optional)
Advanced analysis functions (optional): VCG, timed VCG, QT dispersion, frequency spectrum EKG.
Description
SPECIFICATIONS
Physical Weight & Dimensions
Height: 1.6 inches (40 mm)
Width: 3.9 inches (100 mm)
Depth: 1.6 inches (40 mm)
Weight Approx.: 210 g
General
Connection to PC: USB port
Power supply: DC 5V from USB port
Input circuit: Defibrillation protection built-in
Lead: Standard 12 leads
Acquisition mode: 12 leads simultaneously
Sampling rate: 1000Hz
A/D converter: 12 bits
Input voltage range: ±5mVpp
Time constant: > 3.2s
Frequency response: 0.05Hz ~ 150 Hz
Calibration voltage: 1mV ± 2%
Sensitivity: 2.5, 5, 10, 20 (mm/mV)
Input impedance: 50M Ω (10Hz)
Input circuit current: <50nA
Noise level: <15 μ Vp-p
Baseline drift filter: Baseline forward and backward filter (patent pending)
Patient leakage current: < 10 μ A (220V – 240V)
Patient auxiliary current: <0.1 μ A (DC)
Filter:
EMG filter: 25Hz/35Hz/45Hz/off
Baseline drift filter (patent pending): Strong/weak/off
Lowpass filter: 150 Hz/100 Hz/75 Hz
AC filter: 50 Hz/ 60 Hz
CMRR: >110dB
Rhythm lead: Single lead or three leads selectable
Safety standard: IEC II/CG
PC Requirements
Operating system: Compatible with Windows 8 and Windows 10, Windows 7 OS, Windows 2000 or Windows XP
CPU: Pentium P4, Celeron D 310 or above
RAM: 256 MB or above
HDD: 40G or above
Interface: At least 2 USB ports free